Experience of calibration of a non-contact laser Doppler flowmeter

Year: 2025, volume 21 Issue: №3 Pages: 249-253
Heading: Otorhinolaryngology Article type: Original article
Authors: Mareev G.O., Mareev O.V., Kapralov S.V., Klimenko G.A., Kalyuta T.Yu.
Organization: Saratov State Medical University
Summary:

Objective: to carry out calibration of a non-contact laser Doppler flowmeter (LDF) with justification for its subsequent use in model experiments and clinical trials. Material and methods. An experimental setup is described for reproducing a given blood flow velocity by simulating a capillary; methods are presented for calibrating the flowmeter in tissue perfusion units using a calibration emulsion fluid to determine the instrument readings' dependence on the concentration of moving particles. A method is provided for measuring the level of "biological zero" tissue perfusion of gastrointestinal tract mucosa. Results. Key parameters for processing signals from the non-contact LDF were established along with their relationship to characteristics of the calibration solution's flow and particle movement speed within it. The average value of the "biological zero" index was determined for gastric mucosal wall at 0.07±0.014 tpu. Conclusion. As a result of the carried out work on calibration of the experimental LDF apparatus which measures microcirculatory blood flow using a contactless method, suitability for measurements of the microcirculatory bed of human tissues was confirmed, basic dependences of the parameters of processing the LDF signal on the volume of the flowing calibration emulsion, its speed and the number of moving microparticles in it were obtained.

Bibliography:
1. Holloway GA, Watkins DW. Laser Doppler measurement of cutaneous blood flow. Journal of Investigative Dermatology. 1977;69(3):306-9. DOI:10.1111/1523-1747.ep12507665
2. BonnerRF, Nossal R. Principles of laser-Doppler flowmetry. In: Shepherd A.P., Oberg P.A. (eds). Laser-Doppler Blood Flowmetry. Developments in Cardiovascular Medicine. Boston, MA: Springer, 1990; p. 17-28. DOI:10.1007/978-1-4757-2083-9_2
3. Limjeerajarus С Laser Doppler flowmetry: Basic principle, current clinical and research applications in dentistry. Chulalongkorn University Dental Journal. 2014;37(1): 123-36. DOI:10.58837/CHULA.CUDJ.37.1.11
4. Obeid AN, Barnett NJ, Dougherty G, Ward G. A critical review of laser Doppler flowmetry. J Med Eng Technol. 1990; 14(5): 178-81. DOI: 10.3109/03091909009009955
5. Zacharenko A, Belyaev M, Trushin A, et al. Combined assessment of intestinal viability using laser doppler flowmetry and laser fluorescence spectroscopy. Regional Blood Circulation. DOI:10.24884/1682-6655-2021-20-2-70-76
6. Verhaar N, Geburek F. Real-time ancillary diagnostics for intraoperative assessment of intestinal viability in horses-looking for answers across species. Vet Surg. 2025;54(4):648-64. DOI:10.1111/vsu.14248
7. Verhaar N, Grages A, Bienert-Zeit A, et al. Flowmetry and spectrophotometry for the assessment of intestinal viability in horses with naturally occurring strangulating small intestinal lesions. Equine Vet J. 2024;56(6):1138-48. DOI:10.1111/evj.14118
8. Parshin DS, Topchiyev MA. Peritoneal laser Doppler flowmetry in predicting and diagnosing tertiary peritonitis. Russian Sklifosovsky Journal "Emergency Medical Care". 2020;9(3):410-6. DOI:10.23934/2223-9022-2020-9-3-410-416
9. Adamenkov N, Mamoshin A, Dremin V, et al. Intraoperative assessment of intestinal wall viability (literature review). Journal of Clinical Practice. 2024;15(4):46-58. DOI: 10.17816/clinpract633149
10. Mareev GO. Lazernaya dopplerovskaya floumetriya kak metod otsenki morfofunktsional'nogo sostoyaniya nebnykh mindalin: PhD diss. Saratov, 2004; 156 p.
11. Johnson JM. The cutaneous circulation. In: Laser-Doppler blood flowmetry. Ed. by A.P. Shepherd. KluwerAc. Pub, 1990; p. 121-41.
12. Fairs SLE. Observations of a laser Doppler flowmeter output made using a calibration standard. Med Biol Eng Comput. 1988;26(4):404-6. DOI:10.1007/BF02442299
13. Fredriksson I, Larsson M, Salomonsson F, Stromberg T. Improved calibration procedure for laser Doppler perfusion monitors. 2011. Proc SPIE. 7906. 790602-790602. DOI:10.1117/12.871938
14. Levtov VA, Regirer SA, Shadrina NKh. Reologiya krovi. Moscow: Meditsina, 1982; 270 p.

AttachmentSize
2025_03_249-253.pdf449.97 KB

No votes yet